The flowmeter for highest medium temperatures with genuine loop-powered technology
-
Measuring principle
Electromagnetic
-
Product headline
The Highest Temperature Flowmeter with Real Loop-Powered Technology
Designed for chemical and process applications with corrosive liquids and high medium temperatures.
-
Sensor features
Diverse Applications with Variety of Wetted Materials
Energy-Saving Flow Measurement with No Pressure Loss
Maintenance-Free Design with No Moving Parts
Max. Nominal Diameter: DN 200 (8″)
Common Ex Approvals
Liner Options: PTFE or PFA -
Transmitter features
Convenient Device Wiring with Separate Connection Compartment
Safe Operation with Display, Touch Control, and Background Lighting
Integrated Verification with Heartbeat Technology
Loop-Powered Technology
Robust Dual-Compartment Housing
Worldwide Approvals for Plant Safety (SIL, Hazardous Area) -
Nominal diameter range
DN 15 to 200 (? to 8″)
-
Wetted materials
Liner: PFA; PTFE
Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum -
Measured variables
Volume flow, mass flow
-
Max. measurement error
Volume flow: ±0.5 % o.r. ± 2 mm/s (0.08 in/s)
-
Measuring range
4 dm3/min to 1100 m3/h (1 to 4850 gal/min)
-
Max. process pressure
PN 40, Class 300, 20K
-
Medium temperature range
Liner material PFA: –20 to +150 °C (–4 to +302 °F)
Liner material PTFE: –40 to +130 °C (–40 to +266 °F) -
Ambient temperature range
Flange material carbon steel: –10 to +60 °C (+14 to +140 °F)
Flange material stainless steel: –40 to +60 °C (–40 to +140 °F) -
Sensor housing material
AlSi10Mg, coated
-
Transmitter housing material
AlSi10Mg, coated
-
Degree of protection
IP66/67, type 4X enclosure
-
Display/Operation
4‐line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available -
Outputs
4‐20 mA HART (passive)
Pulse/frequency/switch output (passive) -
Inputs
None
-
Digital communication
HART, PROFIBUS PA, FOUNDATION Fieldbus
-
Power supply
DC 18 to 35 V (4‐20 mA HART with/without pulse/frequency/switch output)
-
Hazardous area approvals
ATEX, IECEx, cCSAus, NEPSI, INMETRO, EAC, JPN, UK Ex
-
Other approvals and certificates
-
Product safety
CE, C-Tick
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (T?V S?D attestation) -
Pressure approvals and certificates
CRN, PED
-
Material certificates
3.1 material
-
Hygienic approvals and certificates
Drinking water approvals: ACS, NSF 61